Features:
| Model | BIOBASE 200 | |
| Sample Pipetting Unit | Sample Rack | 24 sample tubes/rack, at most 23 racks | 
| Primary Sample Tubes | 11.5 – 13mm | |
| Pipetting | 8uL-100uL, (10uL) CV<5% | |
| Time to Dispense | 12 minutes adding 96 samples | |
| Dispensing System | 1 aspirating-dispensing probe with X-Y-Z movement. Dilution system with a syringe of 1000uL, continuous flow washing solution. | |
| Reagent Pipetting Unit | Reagent Racks | 2 reagent tanks (60ml)/rack, at most 23 racks | 
| 
 | QC racks | 24 positions/QC rack, at most 23 racks | 
| 
 | Pipetting | 8uL-100ul, 1ul stepping | 
| 
 | Pipetting Precision | (100ul) CV≥0.5% | 
| 
 | Time to Dispense | 3 minutes (typical)^100uL of 96 reagents to plate from reagent bottle | 
| Washing Unit | Programmable | Yes (options & parameters) | 
| Wash Containers | 3 wash buffers at 2.0L, 1 wash buffers at 5.0L, with level-sensing | |
| Waste Containers | 25L with waste full sensor | |
| Washing System | Two 8 nozzle manifolds, one to dispense and the other one to aspirate | |
| Washing Residual | <2uL | |
| Reading Unit | Reading System | 8 independent photometric channels with mono and bichromatic reading capability | 
| Photometric Range | 0.000-3.000 OD | |
| Spectral Range | 400-700nm | |
| Optical Filters | 2 precision filters (405nm, 630nm) 2 more on request (4 filters max, 405nm, 450nm, 492nm, 630nm) | |
| Reading inaccuracy | Inaccuracy±1% from 0.000 to 1.500 OD, ±2% from 1.500 to 3.000 OD | |
| Incubation Unit | Incubators | 4 independent units | 
| Programmable | Yes | |
| Temperature Range | RT to 45ºC | |
| Temperature Accuracy | ±0.6ºC | |
| Temperature Stability | ±0.6ºC | |